Supportkeyboard_arrow_right

 

No Output
Check
  • Check for miswiring
  • Check for damaged wires
  • Check for leakage to ground
Problem
  • Sensor is damaged or has been damaged
Solution
  • Contact Futek for repair or new sensor
Zero Shift (Less than 10% R.O.)
Check
  • Check for fixture preload
  • Check for miswiring
Problem
  • Sensor has zero shift
Solution
  • Re-balance bridge
    Zero Balance Calculator
Zero Shift (Greater than 10% R.O.)
Check
  • Check for fixture preload
  • Check for miswiring
Problem
  • Sensor may be damaged due to overload
Solution
  • It may be possible to re-balance sensor, but with greatly decreased performance
    Zero Balance Calculator
  • Sensor is irreparably damaged
Incorrect Output
Check
  • Check for improper loading
  • Check for miswiring
  • Check for excessively long cable